Makers of Advanced Rapid Diagnostic Products

Products

LifeSign MI® CK-MB/Myoglobin/Troponin I (3-in-1)

Lifesign mi ckmb 3in1

Immunoassay for the Rapid Qualitative Detection of CK-MB, Myoglobin and
Cardiac Troponin I (cTnI) in Human Whole Blood, Serum and Plasma
as an Aid in the Diagnosis of Myocardial Infarction

  • Easy-to-read color
  • Built-in test control
Clinical Cutoff CK-MB: 5 ng/mL
Myoglobin: 50 ng/mL
Troponin I: 1.5 ng/mL
Time to Results 15 minutes
Specimen Types Whole blood, 120 µL
Serum, 120 µL
Plasma, 120 µL
Internal Control Yes
External Controls Available separately (LSC-301)
Storage 2-30°C (35-86°F)
  CK-MB Myoglobin Troponin I
Comparative Sensitivity 96.7% 95.8% 97.0%
Comparative Specificity 96.9% 98.9% 98.9%
Overall Accuracy 96.8% 98.0% 98.4%
Lifesign mi ckmb myoglobin troponin 3 in 1 procedure
  • Open the foil pouch, remove the LifeSign MI® Myoglobin/CK-MB/Troponin I Test and lay the test on a level surface.
  • Label the test with the patient's identification.
  • Using the dropper provided, add 3 drops (120 µl) of whole blood, serum or plasma into the Sample well.
  • Read the test results at 15 minutes.
Product Kit Size Catalog Number CLIA Complexity
LifeSign MI® CK-MB/Myoglobin/Troponin I (3-in-1) 20 LSC-300-20 Moderate
StatusFirst® Cardiac Control (Level 1/Level 2) 2 x 1 mL LSC-301 N/A